If two instruments read the same sample differently, neither is necessarily faulty – they are most likely measuring two different physical quantities. The catalogue holds five distinct measuring principles; below is where each one gives the right answer.

Why is “method” a question at all?

No portable instrument weighs water directly. Each measures something else that changes with water – resistance, dielectric constant, infrared reflectance – and converts it to a percentage through a calibration curve. The commodity selected on the instrument (wheat, rice or flour) therefore determines the result directly.

The one exception in the catalogue is the loss-on-drying instruments, which really do evaporate the water and weigh the difference.

Electrical resistance

The sample is pressed between two electrodes; resistance falls as water rises. In the catalogue the PRv-930, Riceter FV-202, HB-610, HX-700 and the HK-300 series work on this principle.

Its strength is simplicity and field durability. The HB-610 is the typical example of the class: it measures inside bales and sacks, without disturbing the sample. Its weakness is that resistance is sensitive to moisture distribution and temperature – on a bale with a dry surface and a wet core the reading can mislead.

Capacitance / dielectric

The dielectric constant of water is far above that of grain or wood; the instrument places the sample inside a capacitor and measures that difference. In the catalogue the PM450 and PM650 work at 50 MHz, the HM-540 (wood) and HI-520-2 (concrete and mortar) at 20 MHz.

The advantage is that it sees the whole sample volume and is less distracted by surface moisture than the resistance method. The figure given for the PM650 – “standard error of 0.5% or less against the drying method” – shows the level this class reaches in practice, and at the same time names the reference it is compared against.

NIR – near infrared

Water absorbs specific wavelengths; the instrument measures light reflected from the surface and derives moisture from it. In the catalogue the KB-30 works this way.

It needs no contact, destroys no sample and suits continuous in-line measurement. In return it sees the surface only; colour, particle size and surface texture all affect the result.

Heating and weighing (drying method)

The sample is weighed, dried with heat and weighed again. The difference is water. In the catalogue the FD-660, FD-720 and FD-800 work this way, with ranges given as 0–100% (wet basis) and 0–500% (dry basis).

The reference method is oven drying; the FD series works on the same principle, heating and weighing the sample in a shorter time. The trade-off is that the sample is destroyed. Measure fast on the line, and verify regularly in the laboratory with a drying instrument.

Wet basis or dry basis?

The two separate ranges on the FD series are no accident. Wet basis is water as a share of total weight and cannot exceed 100%. Dry basis is water as a share of dry matter weight and can comfortably pass 100% – which is why the upper limit reads 500%.

For one sample both numbers are correct, but they are not the same number. When buyer and seller speak on different bases, the disagreement comes from the definition.

Instruments in the catalogue and their measuring principles

Ranges and accuracy figures come from the manufacturer data sheets. Accuracy values from different principles cannot be compared directly – they are not defined the same way.

ProductPrincipleRangeSensitivityMaterialPower
Kett PRv-930 Moisture Meter for Flour, Wheat and RiceElectrical resistanceWheat flour 11–18%; Wheat 9–30%; Rice flour 8–18%; Long-grain rice 10–20%; Short-grain rice 10–20%; Long paddy 10–35%; Short paddy 10–30%Flour: 0.25% (SEC, full range); Grain: 0.5% (SEC, below 20%)not stated1.5 V battery × 4
Kett Riceter FV-202 Rice and Grain Moisture MeterElectrical resistancenot stated0.5% (SEC – standard error of the calibration curve, over the 10.0–20.0% range)Long Paddy (LPAD) 10–35%; Long Milled Rice (LMR) 10–20%; Long Brown Rice (LBR) 10–20%; Short Paddy (SPAD) 10–30%; Short Milled Rice (SMR) 10–20%; Short Brown Rice (SBR) 10–20%; Long Parboiled Milled Rice (LPMR) 10–20%; Mixed Parboiled Paddy (MPP) 9–30%1.5 V battery × 4
Kett HB-610 Moisture Meter (Measures Through Bale and Sack)Electrical resistancenot statednot statedRaw cashew / cashew kernel / kernel testa, cocoa bean, green coffee (arabica and robusta), maize, maize flour, wheat, long-grain white rice, long-grain paddy1.5 V AA alkaline battery × 6, approx. 0.45 W
Kett PM450 Grain Moisture MeterCapacitance / dielectric, 50 MHz1–40% depending on the samplenot statedGrain, seed and other product-dependent samples4 × 1.5 V AA alkaline batteries
Kett PM650 Grain Moisture MeterCapacitance / dielectric (50 MHz)1–40% (sample dependent)Standard error of 0.5% or less against the oven-drying method (for all samples below 20% moisture)not stated4 × 1.5 V AA alkaline batteries
Riceter F512 Portable Moisture Meter for Small-Grain Cropsnot stated10–35% (by product)5%not stated4 × AA batteries
PB-3104 Moisture Meter for Flour and Flour ProductsAC resistance format8–20% for all substances±0.1% (production), ±0.5% (oven-drying method)Wheat, flour (pastry & noodles), flour (bread & pasta), noodles, pasta, maize flourAC 220 V or battery (1.5 V alkaline "D" size × 4 – approx. 23 hours continuous operation)
Kett HX-700 Hay Moisture TesterElectrical resistance8–30%Standard error 1.0% (at 20% and below)Hay / straw6 × 1.5 V AA alkaline batteries
HYGROFaster-E Textile Moisture Meternot statedCotton: 0.8–23.3%not statedYarn, fabric, tops, staple fibre, paper and similar materialsnot stated
KB-30 NIR Moisture AnalyserNIR reflectance / absorbancenot statednot statednot statedAC 100–240 V (50/60 Hz) / 50 W
HM-540 Wood Moisture MeterHigh-frequency capacitance (20 MHz)2–150% (depending on specific gravity and wood species)Standard error: 0.5% or less (standard sample, cedar/cypress: 20% or below); oven-drying: JIS Z 2101:2009Wood1 × 9 V alkaline battery (006P), 350 mW
HI-520-2 Concrete and Mortar Moisture MeterHigh-frequency capacitance (20 MHz)not stated±0.5% (concrete, mortar)Lightweight aggregate concrete (LWC/ALA) 0–23%; Plasterboard (GYP) 0–50%; Concrete 0–12%; Autoclaved aerated concrete (ALC) 0–100%; Mortar 0–15%; Calcium silicate board (CSB) 0–15%; D mode 0–1999; S mode (S and D modes only) 0–11001 × 9 V alkaline battery (006P)
HK-300 Series Paper Moisture Meter (-1/-2/-3)Electrical resistanceClamp sensor (HK-300-1): copy paper 4–15%, cardboard 6–20%; Conductive rubber sensor (HK-300-2): copy paper 4–25%, linerboard 6–23%, kraft 5–24%; Constant pressure sensor (HK-300-3): copy paper 2–10%, linerboard 5–15%, kraft 2–10%Standard error 0.4% (at 15% moisture and below)Kraft paper, linerboard, copy paper, cardboard, etc.1.5 V AA alkaline battery × 6, approx. 0.45 W
HX-500 Cotton Moisture Meternot stated4–17% (dry basis)not statedRaw cotton, cotton bale, cotton threadnot stated
FD-720 Infrared Moisture AnalyserLoss on drying (heat-drying and weight-loss method)0–100% (wet basis, solids); 0–500% (dry basis)not statedPowder, granule, liquid, paste, etc.AC 100–120 V / 220–240 V (50/60 Hz)
FD-660 Infrared Moisture AnalyserDetection of weight loss by the heat-drying method0–100% (wet basis, solids) / 0–500% (dry basis)not statednot statedAC 100–120 V / 220–240 V (50/60 Hz)
FD-800 Infrared Moisture AnalyserLoss on drying (heat-drying and weight-loss method)0–100% (wet basis, solids); 0–500% (dry basis)not statedPowder, granule, liquid, paste, etc.AC 100–120 V / 220–240 V (50/60 Hz)
FM-300A Micro Moisture AnalyserRetaining moisture at the surface by filtrationnot statednot statednot statedAC 100–120 V / 220–240 V (50/60 Hz)

Frequently asked questions

  • Which method is the most accurate?

    The reference method is oven drying: the sample is dried and its loss in weight is measured. The FD-660, FD-720 and FD-800 in the catalogue work on the same principle, heating and weighing the sample. The measurement takes longer than other methods and the sample is destroyed.

  • Why do two instruments read the same sample differently?

    Usually because they work on different principles. Resistance is sensitive to moisture distribution, capacitance to sample density, NIR to surface condition. The commodity setting must also be correct on each instrument; measuring flour on a wheat curve produces a systematic offset.

  • Can I measure without destroying the sample?

    Yes. In the catalogue the HB-610 measures inside bales and sacks, and the KB-30 works by contactless NIR. On drying instruments the sample is destroyed by definition.

  • What does a 0–500% range mean?

    It is the dry basis calculation: water is taken as a share of dry matter weight, and that ratio can exceed 100%. On a wet basis the same instrument works over 0–100%.

Not sure which model fits?

Tell us the material, the bag or product size and your daily volume; we will suggest suitable models from the catalogue.