ETV
HYGROFaster-E Textile Moisture Meter
- Sample measuring range
- Cotton: 0.8–23.3%
- Fibre selection
- 24 pure and blended fibres
- Additional measurements
- Ambient relative humidity and temperature
We would like to use a Google Analytics cookie to measure visits. It is not loaded unless you agree; the site works the same either way. Read the privacy notice
KettDistributor in Türkiye
Kett HX-500 cotton moisture meter with four calibration curves for raw cotton, cotton bale and cotton thread; 4–18% measuring range (dry base).

Prices are given in a written quote, by model and quantity.
| Brand | Kett |
|---|---|
| Model | HX-500: cotton moisture meter |
| Measured material | Raw cotton, cotton bale, cotton thread |
| Number of calibration curves | 4 (RAW, BAL, TR1, TR2) |
| Measuring range: raw cotton | 4–17% (dry basis) |
| Measuring range: cotton bale | 4–18% (dry basis) |
| Measuring range: cotton thread | 5–14% (dry basis) |
| Reference method | JIS L 0105: oven drying (105 °C, 2 g, 1 hour) |
| Probe options | Pressure sensor; 21 cm twin-needle probe; 2.5 cm twin-needle probe |
| Calibration accuracy | Correlation R = 0.998–0.999; standard error SE = 0.09–0.27 (by calibration) |
| Calibration table | The table below gives the probe, range and accuracy for all four curves |
Taken from the manufacturer’s catalogue.
Kett HX-500 calibration curves
Scroll sideways to see the whole table.
| Curve | Material | Probe used | Measuring range (dry basis) | Correlation R | Standard error SE | Number of samples |
|---|---|---|---|---|---|---|
| RAW | Raw cotton | Pressure sensor | 4–17% | 0.998 | 0.26 | 11 |
| BAL | Cotton bale | 21 cm twin-needle probe | 4–18% | 0.999 | 0.27 | 6 |
| TR1 | Cotton thread | 2.5 cm twin-needle probe | 5–14% | 0.999 | 0.09 | 6 |
| TR2 | Cotton thread | 21 cm twin-needle probe | 5–14% | — | — | — |
The HX-500 is a hand-held instrument for measuring moisture in cotton processing and textile production. Four calibration curves are selected according to the material: a pressure sensor for raw cotton, a 21 cm twin-needle probe for cotton bales, and a 2.5 cm or 21 cm twin-needle probe for cotton thread. The instrument is calibrated against the JIS L 0105 oven-drying method, so the readings shown are moisture percentages on a dry basis.
Maintenance, repair and spare parts for the machines and instruments we sell are handled by our service workshop at İMÇ, Istanbul. When you ask for a spare part, give the brand, the model and the serial number.
Open the service page
ETV

Dragon

Dragon

Değirmen
Use the “What you need” field to tell us which model and how many units you need. The sales team replies with a written quote covering price, delivery time and service terms.

Use the “Add to quote list” button next to a product or on its page. You can send several products in one request.
Prices are given in a written quote, by model and quantity.